PRODUCTS
ELECTRICAL PROBE SYSTEMS
THERMAL PLATES
LIQUID CRYSTAL SYSTEMS
RESOURCES
SUPPORT
ABOUT
Simple stage for quick experiments with upright and inverted microscopes
-25°C - 90°C with thermoelectric heating and cooling
Open surface with clamps to fit standard 25 mm × 75 mm microscope slides
The TS102S is a Thermoelectric Cooler (Peltier) based thermal stage that can cool down to -25°C without using liquid nitrogen. This is very convenient for performing long duration experiments within a moderate temperature range. The simple system setup reduces experimental complexity compared to LN2 cooled systems or closed chamber stages. With an open surface and spring-loaded clamps to fit 25mm x 75mm microscope slides, this Peltier stage is as quick and easy as things can get for optical experiments requiring temperature control.
Temperature Control |
mK2000 with bidirectional LVDC output |
Thermal Block |
Clear anodized aluminum |
Minimum Temperature |
-10°C continuous operation with C100W, room-temp water -25°C with C100W + Ice water -30°C continuous operation with CW5000 |
Maximum Temperature |
90°C (120ºC upgrade available with customization) |
Temperature Sensor |
100 Ω Platinum RTD |
Maximum Heating Rate |
+25°C per minute at 37°C |
Maximum Cooling Rate |
-15°C per minute at 37°C |
Minimum Heating and Cooling Rate |
±0.01°C per minute |
Temperature Resolution |
0.01°C |
Temperature Stability |
±0.05°C |
Power Supply |
Universal power input - 80W Max (not including water-cooling accessories) |
Software |
Windows software to record and export temperature-time data |
Optical Access |
Reflection and transmission capability |
Minimum Objective Working Distance |
0 mm |
Minimum Condenser Working Distance |
3 mm |
Transmission Aperture |
10 mm diameter, (Options available for Ø5mm, Ø8mm, or Ø12mm) |
Sample Area |
45 mm x 45 mm |
Sample Access |
Open-face with sample clamps and 25mm x 75mm glass slide groove |
Frame Cooling |
Integrated TEC cooling with included C100W water-cooler. Optional chiller upgrade for increased temperature limit is available |
Mounting |
Horizontal or Vertical mounting capability (140mm x 90mm through hole pattern, 99mm diameter dovetail insert) |
Frame Dimensions |
160 mm x 110 mm x 21.7 mm |
Weight |
500g |
TS102S Base-Model Options
Add-on options for the TS102S base model are listed below. These items can be added at any time, and can even be installed by the end-user in the field.
TEC Cooling The standard TS102S comes with a C100W benchtop water circulator that enables sample cooling down to -10°C. By adding ice to the C100W reservoir, the minimum achievable temperature is -25°C, which can be maintained for a short duration until the ice melts. To maintain these lower temperatures for longer intervals, alternative water cooling options are available. The CW5000 is a recirculating chiller uses a compact refrigeration system to cool and circulate water as low as 5°C, allowing the TS102S to maintain -30°C indefinetly.
Note: While operation below 0° is possible with the TS102S, cooling for extended durations in ambient conditions will lead to condensation on the thermal area and sample. A gas-tight or vacuum-tight alternative is reccommended for such applications. | |
- C100W Water Circulator (Standard) Multi-purpose benchtop water circulator PN: C100W |
Upgraded water chiller with refrigeration capabilities, capable of cooling water to 5°C. PN: CW5000 |
Mounting adaptors Mounting adaptors allow the TS102S to be easily mounted on your optical instrument of choice. All TS102S models include 4 through-hold mounting points located at each corner of the frame (140mm x 90mm spacing). The TS102S also includes a ø99mm dove-tail mount integrated into the frame. Besides these two mounting solutions, a number of specially made mounting adaptors are avialable for a wide range of optical tools. | |
- Integrated mounting holes/ dovetail (Standard) PN: N/A |
- Horizontal mounting adaptors |
- Vertical mounting adaptors |
|
|
|
Special options for the TS102S+ customizable model Upgrade to the TS102S+ for access to special options including: | |
Upgraded Temperature Controller Options The TS102S is designed for use with Instec's mK2000 series temperature controllers. For thermoelectric temperature control systems, a bi-directional current output mK2000B is required. Furthermore, the mK2000 itself can be customized with various add-on options such as additional monitoring sensors, analog voltage output and more.... | |
- Additional monitoring sensor for mK2000B PN: mK2000B-TMK |
- Advanced controller upgrades (see mK2000B page for details) Ask sales@instec.com for more details |
Transmission Aperture Some applications require larger transmission apertures than others- whether it be to see more of a sample, allow for more light transmission, or for compatibility with specialized instruments. However, larger transmission apertures also result in larger thermal gradients across the observation area, decreasing the temperature uniformity. Tune the TS102S+ transmission aperture size to optimize the temperature uniformity while maximizing the transmission aperture size. | |
- 10mm (Standard) PN: N/A |
- Reduced 5mm aperture PN: AP5-TS102S |
- Reduced 8mm aperture PN: AP8-TS102S |
- Larger 12mm aperture PN: AP12-TS102S |