Flagship miniature electrical probing stage
-40°C to 150°C* with cryogen-free thermoelectric heating and cooling
4x independent electrical probes with manual positioning
Coaxial/ Triaxial SMA or BNC feedthroughs
40 mm × 40 mm clear anodized aluminum sample area
Sealable chamber with gas purging capability. Optional Vacuum-tight configuration with KF flanges available (TP102V-PM)
Temperature Control |
mK2000B with bidirectional LVDC output |
Minimum Temperature |
-30°C with standard C100W water circulator |
-40°C with upgraded CW5000 chiller | |
Maximum Temperature |
120°C (150°C upgrade available) |
Temperature Sensor |
100 Ω Platinum RTD |
Maximum Heating Rate | +50°C per minute at 37°C |
Maximum Cooling Rate | -18°C per minute at 37°C |
Minimum Heating and Cooling Rate |
±0.01°C per minute |
Temperature Resolution |
0.01°C |
Temperature Stability |
±0.05°C |
Power Supply |
Universal power input - 80W maximum* (* does not include water circluator/ chiller) |
Software |
Windows 10/11 companion software to record and export temperature-time data |
Electrical Probers |
Four (4) DC to 500 MHz electrical probes (200V /1A max) - Prober material options such as tungsten-rhenium, beryllium-copper, or other materials - 25um or 5um tip radius options - Additional standalone electrical feedthroughs available by request |
Probe Positioning |
Manual cantilever probes independently positioned to any point of the sample area before sealing chamber. Ideal for pad sizes larger than 600 um x 600 um* *Smaller pad sizes require -MP or -MPS models |
Connectors |
Coaxial SMA connectors (standard) Coaxial or triaxial BNC options available |
Sample Area Electrical Bias |
Electrically grounded sample area* (standard) (Electrically floating or Triaxial sample area structure available with -MP model) |
Optical Access |
Optical access to sample via reflection (Transmission aperture available with TS102V-PM model) |
Optical Windows |
Removable and exchangeable windows permit full-spectrum transparency |
Minimum Objective Working Distance |
9 mm |
Top Window Aperture |
Ø45 mm aperture (Ø50mm x 1.5mm glass window) |
Top Viewing Angle |
±60° from normal |
Window Defrost |
Optional external window defrosting fixture Note: (fixture included, pressurized air source NOT included) |
Sample Area |
40 mm x 40 mm clear anodized aluminum |
Sample Clips | Optional sample clips - electrically grounded or electrically floating |
Inner Chamber Height |
6.3 mm |
Atmosphere Control |
TP102G-PM - Gas purging capability to control humidity, condensation, and oxidation (maximum pressure +0.5 Bar) TP102V-PM - Includes flexible KF16 vacuum bellows. Supports evacuation down to at least 9.9*10^-3 mBar |
TEC/ Frame Cooling |
Includes water cooling system for TEC, model depends on desired temperature range: -C100W for -30ºC -CW5000 + FVC11 valve box for -40ºC Independent frame cooling loop may be required for certain configurations |
Mounting |
Includes tapped holes on frame and removable side-mounted L-brackets. Horizontal and vertical mounting adaptors available for select instruments, or by custom design |
Chamber Dimensions |
200 mm x 145 mm x 33 mm* *Note: this is the chamber volume ONLY. Overall device footprint may exceed these dimensions. Be sure to consult with sales@instec.com to ensure fitment into optical instruments |
Weight |
1600 g* *Note: weight varies with device configuration |