High temperature thermal stage with LN2 cooling specially design for x-ray measurements. X-ray diffraction and beamline configurations allow for easy use with most systems. Curved kapton window design allows for x-ray measurements over a broad range of diffraction angles.
Model | Unique Features | Additional System Components |
HCP600G-XRD | LN2 cooling with x-ray optical access in a gaspurge chamber |
LN2 Cooling Unit |
HCP421V-XRD | LN2 cooling with x-ray optical access in a vacuum chamber | LN2 Cooling Unit |
HP1200G-XRD | High temperatures with x-ray optical access in gaspurge chamber | Water frame cooling unit |
HP1000V-XRD | High temperatures with x-ray optical access in vacuum chamber | Water frame cooling unit |
Sample Area |
Model Dependent |
Chamber Height |
17 mm |
Sample Access |
Screw-on lid for reliable and easy sample loading |
Atmosphere Control |
Sealed chamber with gas purging capability to control humidity, condensation, oxidation, and vacuum |
Frame Cooling |
Integrated frame cooling with optional chiller system |
Mounting |
Horizontal or Vertical mounting capability |
Frame Dimensions |
150mm x 86 mm x 22 mm |
Weight |
800 g |
Model | Temperature Range | Thermal Area Size | Temperature Sensor |
Temperature Resolution | Thermal Block | Max Heating Rate |
HCP600G-XRD | -196°C to 600°C | Ø26 mm | Pt-RTD | 0.01°C | Silver | +50°C per minute |
HCP421V-XRD | -196°C to 600°C | Ø26 mm | Pt-RTD | 0.01°C | Silver | +50°C per minute |
HP1200G-XRD | 25°C to 1150°C | Ø14 mm | S-type thermocouple | 0.1ºC | Ceramic | +200°C per minute |
HP1000V-XRD | 25°C to 1000°C | Ø14 mm | S-type thermocouple | 0.1ºC | Ceramic | +200°C per minute |