PRODUCTS
ELECTRICAL PROBE SYSTEMS
THERMAL PLATES
LIQUID CRYSTAL SYSTEMS
RESOURCES
SUPPORT
ABOUT
High temperature microscopy & spectroscopy plate
30°C - 1150°C
16mm x 16mm sample area
Gastight chamber*
*Vacuum-tight model available (HP1400V)
Instec's HP1200 Hot Gas-Tight stage can heat samples to 1150°C and are ideal for the study of ceramics, metallurgy, geology and high temperature materials. Electrical feedthroughs and custom sample areas are also available. The HP1200G's gas-tight sample chamber helps to prevent sample oxidation at high temperatures.
Temperature Control |
mK2000 with programmable precision switching PID method |
Thermal Block |
Ceramic |
Sample Thermal Cover |
Removable Inner sample cover with additional window |
Minimum Temperature |
30°C (active cooling by gas purge) |
Maximum Temperature |
1150°C |
Temperature Sensor |
S Type Thermocouple |
Maximum Heating Rate |
+100°C/Min ≤800°C, +20/Min >800°C |
Minimum Heating and Cooling Rate |
±0.5°C per minute |
Temperature Resolution |
0.1°C |
Temperature Stability |
±1°C |
Power supply |
Universal power input |
Software |
Windows compatible software to record and export temperature-time data; remotely control system via VCP connection (Virtual Comm Port) |
Optical access |
Reflection capability (No transmission aperture) |
Optical windows |
Removable and exchangeable windows permit full-spectrum transparency |
Minimum Objective Working Distance |
9.5 mm |
Top window Aperture |
Ø21mm x 0.5mm inner window Ø80 mm x 1mm diameter rotatable outer window |
Top Viewing Angle |
±30° from normal |
Sample Area |
16mm x 16mm (up to 5mm sample height) |
Atmosphere Control |
Gas purging capability to control humidity, condensation, and oxidation |
Frame Cooling |
Integrated frame cooling with optional chiller system |
Window and objective protection |
Cooling by purge air |
Mounting |
Horizontal and vertical mounting |
Frame Dimensions |
200 mm x 120 mm x 30 mm |
Weight |
1350 g |