For IR microscopy & spectroscopy stage
30°C - 1000°C
16 mm X 16mm sample area with wide angle viewing aperture
XY sample positioning
Closed chamber with gas purging capability. Optional Vacuum-tight sample chamber model
Temperature Control |
mK2000 with programmable precision switching PID method |
Thermal Block |
Ceramic |
Sample Thermal Cover |
Removable Inner sample cover with additional window |
Temperature Minimum |
30°C (active cooling not available) |
Temperature Maximum |
1000°C |
Temperature Sensor |
S Type Thermocouple |
Maximum Heating Rate |
+200°C per minute at 100°C |
Minimum Heating and Cooling Rate |
±0.5°C per minute |
Temperature Resolution |
0.1°C |
Temperature Stability |
±1°C |
Power supply |
115 V or 230 V depending on region |
Software |
Windows software to record and export temperature-time data |
Optical access |
Reflection and transmission capability |
Optical windows |
Removable and exchangeable windows permit full-spectrum transparency |
Minimum Objective Working Distance |
10.0 mm |
Minimum Condenser Working Distance |
10.0 mm |
Top Window |
10 mm diameter |
Top Viewing Angle |
±24.5° from normal |
Transmission Aperture |
2.0 mm diameter |
Bottom Viewing Angle |
±10.0° from normal |
Sample Area |
16 mm X 16 mm |
Sample Positioning | 10 mm fine travel with Vernier XY dials for remote manipulation in sealed chambers |
Atmosphere Control |
Gas purging capability to control humidity, condensation, and oxidation |
Frame Cooling |
Integrated frame cooling with optional chiller system |
Window and objective protection |
Cooling by purge air |
Mounting |
Horizontal and vertical mounting capability |
Frame Dimensions |
200 mm x 120 mm x 30 mm |
Weight |
1350 g light-weight aluminum body |