Home | Contact Instec
 
TS102G/V Thermal Gas/Vacuum-Tight Stage
Instec’s TS102 series thermoelectric (Peltier) based thermal vacuum and gas-tight stages provide cooling to -40°C without using liquid nitrogen or a recirculating chiller. This simplifies the system set up, reduces operating costs, and makes it a convenient choice for performing long duration experiments at low temperatures. The TS102V with vacuum chamber capability is ideal for freeze drying samples under a microscope. The TS102G is a lower cost option for controlling the atmosphere when vacuum is not required.
Features
• Programmable precision temperature from -40°C to 120°C Peltier based thermal stage with sealed vacuum or gas-tight sample chamber
• Optional two electric feedthroughs for sample probing
• Controlled fast heating and cooling rate
• Removable and exchangeable windows
• Integrated aperture window defrost system
• Inner cover for improved sample temperature uniformity
• Vertical and horizontal mounting
Technical Specifications
Temperature Range
-40°C to 120°
Temperature Resolution
0.01°C with mK2000
Temperature Stability
±0.05°C at 37°C with mK2000 (better stability optional)
Minimum Heating and Cooling Rate
±0.1°C per hour
Maximum Heating Rate
+80°C per minute at 37°C
Maximum Cooling Rate
-30°C per minute at 37°C
Temperature Control Method
PID with Linear Variable DC
Temperature Control Sensor
100 Ω Platinum RTD
Minimum Objective Working Distance
5.1 mm
Minimum Condenser Working Distance
17.5 mm
Sample Area
25 mm x 75 mm
Chamber Height
3.5 mm
Sample Viewing Aperture
2 mm for transmitted light
26.5 mm (G) / 38.5 mm (V) for reflected light
Vacuum Port
KF25
Weight
1,150 g (G) / 1,210 g (V)
 
*Sample area, chamber height and sample viewing aperture specifications given without inner cover in place. 5 or 8 mm aperture for transmitted light available by request.
 